4 Use
The measurement boxes also serve to set the alarm thresholds.
4.6.17.1 Adjustment of alarm thresholds
The alarm thresholds can be accessed and adjusted directly via the ventilation screen.
To adjust a threshold:
•
Select the high or low threshold: it is highlighted.
•
Adjust the value by turning the rotating knob and then press the rotating knob to validate.
When an alarm threshold is exceeded, the box and the corresponding alarm threshold turn red. This
display remains for as long as the alarm is present.
When the alarm disappears, the box returns to its original state but the threshold concerned remains in
red to indicate that an alarm has been activated.
Press the threshold to end the display of the threshold in red.
CAUTION:
After restarting the unit or changing the patient category, the alarm thresholds are
automatically reset to their standard value.
CAUTION:
It is necessary to set these alarm thresholds to ensure that they are suitable for
the patient and his/her ventilation. For example: the high-pressure threshold is very important
to protect the patient from pressure levels that are too high in the airways.
CAUTION:
The setting of alarm thresholds to the extreme values in the adjustment ranges
can render the alarm system ineffective.
57