9 Technical description
Immunity test
Test level
Compliance level
Electromagnetic environ-
ment/comments
Conducted disturbances, in-
duced by radio-frequency
fields (IEC 610004-6)
3 V
150KHz to 80MHz
6 V ISM bandwidth and band-
widths between 0.15 MHZ
and 80 MHZ, including ama-
teur radio bandwidths
80% MA to 1 KHz
3 V
150KHz to 80MHz
6 V ISM bandwidth and band-
widths between 0.15 MHZ
and 80 MHZ, including ama-
teur radio bandwidths
80% MA to 1 KHz
The electromagnetic field strengths of fixed radio-frequency transmitters, as determined by a survey of the electromagnetic
environment (a), must be below the compliance level for each frequency range. Interference may occur close to equipment
identified by the following symbol:
Comment: These specifications may not apply in all situations. Electromagnetic propagation is affect-
ed by absorption and reflection from structures, objects and people.
(a) The electromagnetic field strengths of fixed radio-frequency transmitters, such as base stations for
mobile phones (cellular/wireless), mobile radios, amateur radio operators, AM/FM radio broadcasts
and TV broadcasts cannot be determined accurately through theory. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be carried out. If the
intensity level of the radio-frequency field in the immediate environment in which the product is being
used exceeds the radio-frequency compliance level specified above, the product performance must be
tested in order to check whether or not it is compliant with the specifications. In the case of any abnor-
mal performance of the device, additional measures may be necessary, such as reorienting or relocat-
ing the product.