Monnal T60 Advanced - User manual
Use with a defibrillator
•
When using the
Monnal T60 Advanced
and a defibrillator simultaneously, the defibrillation shock
in the presence of enriched oxygen and combustibles (such as textiles) can cause an explosion or
fire which could injure the patient and bystanders.
•
It is recommended to use adhesive electrodes.
•
During the defibrillation:
•
remove the oxygen mask or the nasal cannula and keep it at least 1 m from the patient’s torso;
•
if the patient is intubated, leave the ventilator connected;
•
ensure that the oxygen-enriched air at the outlet of the expiratory valve is not facing the patient’s
torso.
Electrical power supply
•
Check that the voltage in the mains socket used matches the electrical characteristics of the venti-
lator (indicated on the rear panel of the power supply adapter).
•
Use only the mains cable and mains power supply box supplied with the device.
•
If an external power supply is used, check that the voltage and current used match the electrical
characteristics of the ventilator (indicated on the side of the ventilator).
•
The mains power supply adapter is not protected from splashes of water (IPX0), unlike the device
itself, which complies with IPX4 during battery-powered operation.
•
This ventilator is equipped with an internal battery and a removable battery.
•
The device must be connected to the mains regularly to maintain the battery charge at a suitable
level.
•
In the event of any doubt about the condition of the mains power supply cable, use the device on
its internal battery only.
•
If used for long periods with the battery, it is recommended that an additional extractable battery
is kept at hand.
•
Do not use antistatic or electrically conductive pipes.
•
The user must not touch the patient and the equipment enclosures at the same time.
Internal-battery power supply
The internal battery is only a backup power supply. If the device cannot be connected to an external
power source (primary source), an extractable battery (secondary source) should be used. Using the
internal battery as the main power source can cause ventilation to stop unexpectedly.
IP Protection
To ensure the IP protection level of the device is maintained during normal use, it is essential that all
removable components (air filter, expiratory assembly, O2 sensor cover and the rear plastic panel) are
fitted in place.
Electromagnetic compatibility
•
The presence of equipment as diathermy units, high frequency electro-surgical equipment, defibril-
lators and cellular telephones or of electromagnetic interferences exceeding IEC 60601-1-2 levels
in its vicinity may interfere with the normal operation of the ventilator.
•
The
Monnal T60 Advanced
should not be placed next to or on top of this equipment. If such use is
necessary, the
Monnal T60 Advanced
should be observed together with the other ventilators to en-
sure that they are operating normally.
•
Do not use this ventilator in a specifically magnetic environment (MRI, NMR, etc.).
•
Monnal T60 Advanced
is compliant with the requirements defined in standard EN 60601-1-2 relat-
ing to the electromagnetic compatibility of medical devices. Precautionary measures are required
with this device in terms of EMC and the devices must be installed and put into operation in accor-
dance with the EMC information provided in this user manual.