Connect the expiratory branch to the expiratory valve of the ventilator:
(27)
•
Connect the inspiratory branch to the inspiratory outlet cone of the ventilator: (28)
CAUTION:
Make sure to limit dead space when installing the patient circuit and accessories.
CAUTION:
When the machine is used on a new patient, it is necessary to follow the hospital's
hygiene protocol for new single-use material and for the disinfection of reusable material,
as appropriate. The accessories and consumables (patient circuit, masks, expiratory valves,
adapters, nebulizer, etc.) are often available in a single-use and an autoclavable version.
CAUTION:
Reusing single-use accessories or consumables carries the risk of patient cross-
contamination. This risk also arises if reusable accessories or consumables are not sterilized
between each use.
3.7 CO
2
Measurement Probe (IRMA
™
)
Note:
CO
2
monitoring requires a software option that can be enabled using a code. To use
this option, contact your Air Liquide Medical Systems representative
After purchasing this option,
Air Liquide Medical Systems
provides:
•
the activation code for the option,
•
the CO2 probe for measuring the concentration of exhaled carbon dioxide (in accordance with ISO
80601-2-55),
•
the necessary adapters.
1.
Connect the IRMA
™
probe to the etCO2 connection socket.
22