5 Alarms and other messages
No.
Alarms
Priority
Alarm activation
Alarm activation delay
Actions
8
Patient disconnection!!!
HIGH
Patient disconnection
or excessive leakage, or
electronic malfunction
Inhibition for 60 s when
ventilation starts.
Inhibition during a
pause and 30 s after the
pause.
Otherwise triggering
within 17 s.
Check the leakage level.
Check the "adult/child"
patient setting.
Ensure the ventilation
settings are suitable for
the patient.
Contact the technical
department if the prob-
lem continues.
9
High pressure!!!
HIGH
Peak pressure above
alarm threshold
Triggered after
3 consecutive
ventilation cycles with a
Pmax
pressure above the
alarm limit
Check the coherence of
alarm levels with venti-
lation settings
When the pressure
threshold is reached,
the machine switches to
the expiration phase.
10
Error detected!!! Contact
the technical depart-
ment
HIGH
Electronics failure
Connectivity tests per-
formed every second
Use a different unit and
contact your technical
department
11
Error detected!!! Contact
the technical depart-
ment
HIGH
Electronics failure
Immediate triggering
Use a different unit and
contact your technical
department
12
Screen fault!!! Ventila-
tion Effective
HIGH
Electronics failure
Immediate triggering
Use a different unit and
contact your technical
department
13
Incorrect Settings!!!
HIGH
Electronics failure
Immediate triggering
Use a different unit and
contact your technical
department
Ventilation continues
in VCV mode with the
default settings.
14
Incorrect Settings!!!
HIGH
Ventilation settings are
not technically achiev-
able, or a software fault
has occurred
Immediate triggering
Check and amend the
settings if necessary.
Contact your technical
department if the prob-
lem persists or is recur-
rent.
Ventilation continues
but the new settings are
not adopted.
83