Monnal T60 Advanced - User manual
Note:
The audible alarm is for a health professional near the patient. The maximum distance
away and the volume of the alarm must therefore be determined by the user according to the
situation.
CAUTION:
You must set a higher volume than the ambient noise. A volume that is too low
may prevent users from hearing the alarm.
Brightness
To adjust the brightness of the screen backlighting:
•
Press the
Brightness
key,
•
Turn the control wheel to display the desired value,
•
Press the control wheel again to validate.
Keypad tone
The
Keypad tone
function allows the user to hear a beeping sound whenever a key on the screen is
pressed:
1.
Press the
Keypad tone
to activate the function (by default, this function is deactivated).
2.
Press the
Keypad tone
again to deactivate the function.
Alarm flash
The
Alarm flash
function allows you to enable the presence of the casing indicator light on the front
panel of the machine.
•
Press the
Alarm flash
key to activate the function. This function is enabled by default.
•
Press the
Alarm flash
key again to disable the function.
Date / Time (only on standby)
To set the date and time on the machine, press the
Date / Time
key.
To set each parameter (day / month / year + hour / minute):
1.
Turn the rotating knob to display the desired value,
2.
Press the rotating knob to validate and move to the next value.
4.7.5 Low-pressure O
2
This function allows the machine to enrich the O
2
mixture from a low-pressure source, typically a low
flow meter on an O
2
cylinder or a concentrator.
This is different from connecting it to a high-pressure source, as it involves lower quantities (pressure
generally below 1 bar, flow less than 10 L/min); According to the nature of the source and the ventila-
tion settings, certain FiO2 values cannot be guaranteed. An alignment chart is provided for information
in Section Technical description on page 107.
To enable the low-pressure O
2
function, press
Menu
and then
Low-pressure O
2
.
The message
‘Low-pressure O
2
enabled’
(information) then appears in the alarm panel.
The gas mixture is O
2
-enriched and monitored in conjunction with the FiO
2
setting parameter, and the
machine supplies the desired concentration insofar as possible.
If the ventilation and low-pressure source settings are incompatible, the FiO
2
will not be reached, and
the
Low FiO
2
!!!
alarm will be triggered. The user will then be informed that the settings need to be mod-
ified.
68
Summary of Contents for MONNAL T60 ADVANCED