Monnal T60 Advanced - User manual
No.
Alarms
Priority
Alarm activation
Alarm activation delay
Actions
67
Ventilator operating on
internal battery!!
MEDIUM
The machine is operat-
ing on internal battery
Immediate
triggering
Replace the auxiliary
battery or connect the
machine to AC mains.
69
FiO2 sensor requires re-
placement soon!
LOW
O2 sensor at end of life
Immediate
triggering
Press the control wheel
to acknowledge the
alarm.
Replace the O2 sensor
before you next put the
unit in use.
70
Non Critical error! Venti-
lation Effective
LOW
Electronics failure
Triggering during auto-
matic tests
Re-run the automatic
tests.
If the problem persists:
Use a different unit and
contact your technical
department
71
Screen locked
None
(INFO cat-
egory)
Pressing the screen
when it is locked
Immediate
triggering
Follow the on-screen in-
structions to unlock if
necessary
72
Unit on stand-by
None
(INFO cat-
egory)
Unit on stand-by
Immediate triggering
when ventilation shuts
down or when unit is
started up
Not applicable.
73
Screen upside down
None
(INFO cat-
egory)
Press the 180° key
Immediate triggering
Press the control wheel
to confirm screen inver-
sion.
Press ‘cancel’ to leave
the screen in the same
orientation.
74
Ventilation Interrupted!!!
use an alternative venti-
lator
HIGH (UL-
TRA cate-
gory)
Turbine
temperature too high
Immediate triggering
Check the machine air
inlet.
Replace the filters.
Use a different unit and
contact your technical
department
75
Non Critical error! Venti-
lation Effective
LOW
Electronics failure
Immediate triggering
Use a different unit and
contact the technical
department
76
CO2 apnea!!!
MEDIUM
No patient exhalation
gas been detected for
more than 20 s.
Immediate triggering
A resuscitation special-
ist must re-evaluate the
patient’s ventilation.
77
Check the adapter of
the IRMA™ (CO2) probe
MEDIUM
UU adapter poorly fitted
on the probe (The red
LED on the probe flash-
es)
Immediate triggering
Check the UU adapter.
78
Replace the adapter of
the IRMA™ (CO2) probe
MEDIUM
UU adapter obstructed
or blocked
Immediate triggering
Change the adapter.
90