4 Use
4.6.17.3 Display of measurements
List of measured parameters (those with alarm thresholds must be present) on two pages:
Measurements
Type of information
FiO2
Insufflated oxygen fraction measured every second
VTe
Volume exhaled during the cycle
RR
Respiratory rate averaged over 4 cycles
Ti/Ttot
Ratio of inspiratory time to total time
I:E
Ratio of inspiration time to expiration time
Ppeak
Peak inspiratory pressure of the cycle
Pplat
Pressure measured at the end of the plateau or inspiratory pause
PEEP
Pressure measured 80 ms before the end of expiration
VTi
Volume insufflated during the cycle (invasive ventilation)
MVe
Expiratory volume per minute averaged over 30 s
MVi
Inspiratory volume per minute averaged over 30 s.
Leak
Leak percentage (non-invasive ventilation)
Pmean
Mean pressure of respiratory cycle
etCO2
Expired CO2 fraction at end of expiration
CO2
Highest instantaneous CO2 value between two ventilation cycles.
% CC
Duration of cardiac massage in relation to time spent in the chest compression phase (%)
Timer
Time since CPV mode started
.
P-P
Maximum pressure variation generated by the chest compressions (cmH2O)
Vt divided by pre-
dicted weight (Vt/
PBW)
Ratio of the tidal volume to the predicted weight
.
spont MVe
Spontaneous exhaled minute volume
.
spont RR
Spontaneous respiratory rate
.
Driving pressure
(DP)
Plateau pressure - PEEP
.
Cstat
Measurement of static compliance
. See Cstat measurement on page 63.
DynC
Measurement of estimated compliance
.
2
If the etCO2 software option is enabled.
59