514
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
23
:MTESt Commands
:MTESt:AMASk:SOURce
(see
page 1208
)
Command Syntax
:MTESt:AMASk:SOURce <source>
<source> ::= CHANnel<n>
<n> ::= 1 to (# analog channels) in NR1 format
The :MTESt:AMASk:SOURce command selects the source for the
interpretation of the :MTESt:AMASk:XDELta and :MTESt:AMASk:YDELta
parameters when :MTESt:AMASk:UNITs is set to CURRent.
When UNITs are CURRent, the XDELta and YDELta parameters are
defined in terms of the channel units, as set by the :CHANnel<n>:UNITs
command, of the selected source.
Suppose that UNITs are CURRent and that you set SOURce to CHANNEL1,
which is using units of volts. Then you can define AMASk:XDELta in
terms of volts and AMASk:YDELta in terms of seconds.
This command is the same as the :MTESt:SOURce command.
Query Syntax
:MTESt:AMASk:SOURce?
The :MTESt:AMASk:SOURce? query returns the currently set source.
Return Format
<source> ::= CHAN<n>
<n> ::= 1 to (# analog channels) in NR1 format
See Also
•
"Introduction to :MTESt Commands"
on page 509
•
":MTESt:AMASk:XDELta"
on page 516
•
":MTESt:AMASk:YDELta"
on page 517
•
":MTESt:AMASk:UNITs"
on page 515
•
":MTESt:SOURce"
on page 539
Example Code
•
"Example Code"
on page 509
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...