552
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
25
:POWer Commands
:POWer:DESKew
(see
page 1208
)
Command Syntax
:POWer:DESKew
The :POWer:DESKew command launches the auto deskew process on the
oscilloscope.
Before sending this command:
1
Demagnetize and zero- adjust the current probe.
Refer to the current probe's documentation for instructions on how to
do this.
2
Make connections to the U1880A deskew fixture as described in the
oscilloscope's connection dialog or in the
DSOX4PWR Power
Measurement Application User's Guide
.
3
Make sure the voltage probe and current probe channels are specified
appropriately using the :POWer:SIGNals:SOURce:VOLTage1 and
:POWer:SIGNals:SOURce:CURRent1 commands.
The deskew values are saved in the oscilloscope until a factory default or
secure erase is performed. The next time you run the Power Application,
you can use the saved deskew values or perform the deskew again.
Generally, you need to perform the deskew again when part of the test
setup changes (for example, a different probe, different oscilloscope
channel, etc.) or when the ambient temperature has changed.
See Also
•
":POWer:SIGNals:SOURce:VOLTage<i>"
on page 599
•
":POWer:SIGNals:SOURce:CURRent<i>"
on page 598
N O T E
Use the lowest attenuation setting on the high voltage differential probes whenever
possible because the voltage levels on the deskew fixture are very small. Using a higher
attenuation setting may yield inaccurate skew values (and affect the measurements made)
because the noise level is magnified as well.
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...