1154
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
36
Obsolete and Discontinued Commands
:MTESt:RUMode
(see
page 1208
)
Command Syntax
:MTESt:RUMode {FORever | TIME,<seconds> | {WAVeforms,<wfm_count>}}
<seconds> ::= from 1 to 86400 in NR3 format
<wfm_count> ::= number of waveforms in NR1 format
from 1 to 1,000,000,000
The :MTESt:RUMode command determines the termination conditions for
the mask test. The choices are FORever, TIME, or WAVeforms.
•
FORever — runs the Mask Test until the test is turned off.
•
TIME — sets the amount of time in seconds that a mask test will run
before it terminates. The <seconds> parameter is a real number from 1
to 86400 seconds.
•
WAVeforms — sets the maximum number of waveforms that are required
before the mask test terminates. The <wfm_count> parameter indicates
the number of waveforms that are to be acquired; it is an integer from
1 to 1,000,000,000.
Query Syntax
:MTESt:RUMode?
The :MTESt:RUMode? query returns the currently selected termination
condition and value.
Return Format
{FOR | TIME,<seconds> | {WAV,<wfm_count>}}<NL>
<seconds> ::= from 1 to 86400 in NR3 format
<wfm_count> ::= number of waveforms in NR1 format
from 1 to 1,000,000,000
See Also
•
"Introduction to :MTESt Commands"
on page 509
•
":MTESt:RUMode:SOFailure"
on page 1155
N O T E
The :MTESt:RUMode command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MTESt:RMODe command (see
page 526
) instead.
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...