Root (:) Commands
6
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
199
:DIGitize
(see
page 1208
)
Command Syntax
:DIGitize [<source>[,..,<source>]]
<source> ::= {CHANnel<n> | FUNCtion<m> | MATH<m> | SBUS{1 | 2}}
for the DSO models
<source> ::= {CHANnel<n> | DIGital<d> | POD{1 | 2}
| BUS{1 | 2} | FUNCtion<m> | MATH<m> | SBUS{1 | 2}}
for the MSO models
<n> ::= 1 to (# analog channels) in NR1 format
<d> ::= 0 to (# digital channels - 1) in NR1 format
<m> ::= 1-4 in NR1 format
The <source> parameter may be repeated up to 5 times.
The :DIGitize command is a specialized RUN command. It causes the
instrument to acquire waveforms according to the settings of the :ACQuire
commands subsystem. When the acquisition is complete, the instrument is
stopped.
If no argument is given, :DIGitize acquires the channels currently
displayed. If no channels are displayed, all channels are acquired.
See Also
•
"Introduction to Root (:) Commands"
on page 190
•
":RUN"
on page 220
•
":SINGle"
on page 222
•
":STOP"
on page 224
•
":TIMebase:MODE"
on page 925
•
Chapter 7
, “:ACQuire Commands,” starting on page 227
•
Chapter 33
, “:WAVeform Commands,” starting on page 1019
N O T E
The :DIGitize command is only executed when the :TIMebase:MODE is MAIN or WINDow.
N O T E
To halt a :DIGitize in progress, use the device clear command.
N O T E
MATH<m> is an alias for FUNCtion<m>.
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...