:WAVeform Commands
33
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
1021
Introduction to
:WAVeform
Commands
The WAVeform subsystem is used to transfer data to a controller from the
oscilloscope waveform memories. The queries in this subsystem will only
operate when the channel selected by :WAVeform:SOURce is on.
Waveform Data and Preamble
The waveform record is actually contained in two portions: the preamble
and waveform data. The waveform record must be read from the
oscilloscope by the controller using two separate commands,
:WAVeform:DATA (see
page 1029
) and :WAVeform:PREamble (see
page 1036
). The waveform data is the actual data acquired for each point
in the specified source. The preamble contains the information for
interpreting the waveform data, which includes the number of points
acquired, the format of acquired data, and the type of acquired data. The
preamble also contains the X and Y increments, origins, and references for
the acquired data, so that word and byte data can be translated to time
and voltage values.
:WAVeform:UNSigned
{{0 | OFF} | {1 |
ON}} (see
page 1047
)
:WAVeform:UNSigned?
(see
page 1047
)
{0 | 1}
:WAVeform:VIEW <view>
(see
page 1048
)
:WAVeform:VIEW? (see
page 1048
)
<view> ::= {MAIN}
n/a
:WAVeform:XINCrement?
(see
page 1049
)
<return_value> ::= x-increment
in the current preamble in NR3
format
n/a
:WAVeform:XORigin?
(see
page 1050
)
<return_value> ::= x-origin
value in the current preamble in
NR3 format
n/a
:WAVeform:XREFerence?
(see
page 1051
)
<return_value> ::= 0
(x-reference value in the current
preamble in NR1 format)
n/a
:WAVeform:YINCrement?
(see
page 1052
)
<return_value> ::= y-increment
value in the current preamble in
NR3 format
n/a
:WAVeform:YORigin?
(see
page 1053
)
<return_value> ::= y-origin in
the current preamble in NR3
format
n/a
:WAVeform:YREFerence?
(see
page 1054
)
<return_value> ::= y-reference
value in the current preamble in
NR1 format
Table 144 :WAVeform Commands Summary (continued)
Command
Query
Options and Query Returns
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...