936
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
32
:TRIGger Commands
The following trigger types are available (see
":TRIGger:MODE"
on
page 944).
• Edge triggering
— identifies a trigger by looking for a specified slope
and voltage level on a waveform.
• Nth Edge Burst triggering
— lets you trigger on the Nth edge of a burst
that occurs after an idle time.
• Pulse width triggering
— (:TRIGger:GLITch commands) sets the
oscilloscope to trigger on a positive pulse or on a negative pulse of a
specified width.
• Pattern triggering
— identifies a trigger condition by looking for a
specified pattern. This pattern is a logical AND combination of the
channels. You can also trigger on a specified time duration of a pattern.
• TV triggering
— is used to capture the complicated waveforms of
television equipment. The trigger circuitry detects the vertical and
horizontal interval of the waveform and produces triggers based on the
TV trigger settings you selected. TV triggering requires greater than º
division of sync amplitude with any analog channel as the trigger
source.
• USB (Universal Serial Bus) triggering
— will trigger on a Start of
Packet (SOP), End of Packet (EOP), Reset Complete, Enter Suspend, or
Exit Suspend signal on the differential USB data lines. USB Low Speed
and Full Speed are supported by this trigger.
Reporting the Setup
Use :TRIGger? to query setup information for the TRIGger subsystem.
Return Format
The return format for the TRIGger? query varies depending on the current
mode. The following is a sample response from the :TRIGger? query. In
this case, the query was issued following a *RST command.
:TRIG:MODE EDGE;SWE AUTO;NREJ 0;HFR 0;HOLD +60.0000000000000E-09;
:TRIG:EDGE:SOUR CHAN1;LEV +0.00;SLOP POS;REJ OFF;COUP DC;
:TRIG:ZONE:STAT 0
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...