:COMPliance Commands
11
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
291
:COMPliance:USB:TEST:CONNection
(see
page 1208
)
Command Syntax
:COMPliance:USB:TEST:CONNection <connection>
<connection> ::= {SINGleended | DIFFerential}
When a Hi- Speed test has been selected, the
:COMPliance:USB:TEST:CONNection command specifies the test fixture
connection type:
•
DIFFerential — specifies that a differential probe is used to probe the
signal under test.
In this case, use the :COMPliance:USB:SOURce:DIFFerential command to
specify the analog input channel connected to the differential probe.
•
SINGleended — specifies that two single- ended SMA cables are used to
probe the signal under test.
In this case, use the :COMPliance:USB:SOURce:DPLus command to
specify the analog input channel that is probing the D+ signal, and use
the :COMPliance:USB:SOURce:DMINus command to specify the analog
input channel that is probing the D- signal.
Query Syntax
:COMPliance:USB:TEST:CONNection?
The :COMPliance:USB:TEST:CONNection? query returns the specified the
test fixture connection type.
Return Format
<connection><NL>
<connection> ::= {SING | DIFF}
See Also
•
":COMPliance:USB:SOURce:DIFFerential"
on page 287
•
":COMPliance:USB:SOURce:DPLus"
on page 289
•
":COMPliance:USB:SOURce:DMINus"
on page 288
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...