s
1
227
Agilent InfiniiVision 4000 X-Series Oscilloscopes
Programmer's Guide
7
:ACQuire Commands
Set the parameters for acquiring and storing data. See
"Introduction to
:ACQuire Commands"
on page 227.
Introduction to
:ACQuire
Commands
The ACQuire subsystem controls the way in which waveforms are
acquired. These acquisition types are available: normal, averaging, peak
detect, and high resolution.
Table 82 :ACQuire Commands Summary
Command
Query
Options and Query Returns
:ACQuire:COMPlete
<complete> (see
page 229
)
:ACQuire:COMPlete?
(see
page 229
)
<complete> ::= 100; an integer in
NR1 format
:ACQuire:COUNt
<count> (see
page 230
)
:ACQuire:COUNt? (see
page 230
)
<count> ::= an integer from 2 to
65536 in NR1 format
:ACQuire:MODE <mode>
(see
page 231
)
:ACQuire:MODE? (see
page 231
)
<mode> ::= {RTIMe | ETIMe |
SEGMented}
n/a
:ACQuire:POINts? (see
page 232
)
<# points> ::= an integer in NR1
format
:ACQuire:RSIGnal
<ref_signal_mode>
(see
page 233
)
:ACQuire:RSIGnal?
(see
page 233
)
<ref_signal_mode> ::= {OFF | OUT
| IN}
:ACQuire:SEGMented:AN
ALyze (see
page 234
)
n/a
n/a (with Option SGM)
:ACQuire:SEGMented:CO
UNt <count> (see
page 235
)
:ACQuire:SEGMented:CO
UNt? (see
page 235
)
<count> ::= an integer from 2 to
1000 in NR1 format (with Option
SGM)
:ACQuire:SEGMented:IN
Dex <index> (see
page 236
)
:ACQuire:SEGMented:IN
Dex? (see
page 236
)
<index> ::= an integer from 1 to
1000 in NR1 format (with Option
SGM)
n/a
:ACQuire:SRATe? (see
page 239
)
<sample_rate> ::= sample rate
(samples/s) in NR3 format
:ACQuire:TYPE <type>
(see
page 240
)
:ACQuire:TYPE? (see
page 240
)
<type> ::= {NORMal | AVERage |
HRESolution | PEAK}
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...