Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
33
:WAVeform Commands
:WAVeform:POINts:MODE
<points_mode> (see
page 1034
)
:WAVeform:POINts:MODE
? (see
page 1034
)
<points_mode> ::= {NORMal |
MAXimum | RAW}
n/a
:WAVeform:PREamble?
(see
page 1036
)
<preamble_block> ::= <format
NR1>, <type NR1>,<points
NR1>,<count NR1>, <xincrement
NR3>, <xorigin NR3>, <xreference
NR1>,<yincrement NR3>, <yorigin
NR3>, <yreference NR1>
<format> ::= an integer in NR1
format:
• 0 for BYTE format
• 1 for WORD format
• 2 for ASCii format
<type> ::= an integer in NR1
format:
• 0 for NORMal type
• 1 for PEAK detect type
• 3 for AVERage type
• 4 for HRESolution type
<count> ::= Average count, or 1
if PEAK detect type or NORMal; an
integer in NR1 format
n/a
:WAVeform:SEGMented:C
OUNt? (see
page 1039
)
<count> ::= an integer from 2 to
1000 in NR1 format (with Option
SGM)
n/a
:WAVeform:SEGMented:T
TAG? (see
page 1040
)
<time_tag> ::= in NR3 format
(with Option SGM)
:WAVeform:SOURce
<source> (see
page 1041
)
:WAVeform:SOURce?
(see
page 1041
)
<source> ::= {CHANnel<n> |
FUNCtion<m> | MATH<m> | SBUS} for
DSO models
<source> ::= {CHANnel<n> | POD{1
| 2} | BUS{1 | 2} | FUNCtion<m> |
MATH<m> | SBUS} for MSO models
<n> ::= 1 to (# analog channels)
in NR1 format
<m> ::= 1-4 in NR1 format
:WAVeform:SOURce:SUBS
ource <subsource>
(see
page 1045
)
:WAVeform:SOURce:SUBS
ource? (see
page 1045
)
<subsource> ::= {{SUB0 | RX |
MOSI} | {SUB1 | TX | MISO}}
n/a
:WAVeform:TYPE? (see
page 1046
)
<return_mode> ::= {NORM | PEAK |
AVER | HRES}
Table 144 :WAVeform Commands Summary (continued)
Command
Query
Options and Query Returns
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...