324
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
15
:DVM Commands
:DVM:ARANge
(see
page 1208
)
Command Syntax
:DVM:ARANge <setting>
<setting> ::= {{OFF | 0} | {ON | 1}}
If the selected digital voltmeter (DVM) source channel is not used in
oscilloscope triggering, the :DVM:ARANge command turns the digital
voltmeter's Auto Range capability on or off.
•
When on, the DVM channel's vertical scale, vertical (ground level)
position, and trigger (threshold voltage) level (used for the counter
frequency measurement) are automatically adjusted.
The Auto Range capability overrides attempted adjustments of the
channel's vertical scale and position.
•
When off, you can adjust the channel's vertical scale and position
normally.
Query Syntax
:DVM:ARANge?
The :DVM:ARANge? query returns a flag indicating whether the digital
voltmeter's Auto Range capability is on or off.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also
•
":DVM:SOURce"
on page 329
•
":DVM:ENABle"
on page 326
•
":DVM:MODE"
on page 328
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...