:ACQuire Commands
7
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
233
:ACQuire:RSIGnal
(see
page 1208
)
Command Syntax
:ACQuire:RSIGnal <ref_signal_mode>
<ref_signal_mode> ::= {OFF | OUT | IN}
The :ACQuire:RSIGnal command selects the 10 MHz reference signal mode.
•
The OFF mode disables the oscilloscope's 10 MHz REF BNC connector.
•
The OUT mode is used to synchronize the timebase of two or more
instruments.
•
The IN mode is used to supply a sample clock to the oscilloscope. A
10 MHz square or sine wave signal is input to the BNC connector
labeled 10 MHz REF. The amplitude must be from - 5 dBm to 17 dBm
(356 mVpp to 4.48 Vpp).
Query Syntax
:ACQuire:RSIGnal?
The :ACQuire:RSIGnal? query returns the current 10 MHz reference signal
mode.
Return Format
<ref_signal_mode><NL>
<ref_signal_mode> ::= {OFF | OUT | IN}
See Also
•
":TIMebase:REFClock"
on page 928
•
The
Agilent InfiniiVision 4000 X Series Oscilloscopes User's Guide
for
information on using the 10 MHz reference clock.
C A U T I O N
Do not apply more than 20 dBm Max (6.32 Vpp Max) at the 10 MHz REF BNC
connector on the rear panel or damage to the instrument may occur.
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...