334
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
16
:EXTernal Trigger Commands
:EXTernal:RANGe
(see
page 1208
)
Command Syntax
:EXTernal:RANGe <range>[<suffix>]
<range> ::= vertical full-scale range value in NR3 format
<suffix> ::= {V | mV}
The :EXTernal:RANGe command is provided for product compatibility.
When using 1:1 probe attenuation, the range is either 1.6 V or 8 V.
The range is automatically recalculated when the external trigger probe
attenuation factor is changed.
Query Syntax
:EXTernal:RANGe?
The :EXTernal:RANGe? query returns the current full- scale range setting
for the external trigger.
Return Format
<range_argument><NL>
<range_argument> ::= external trigger range value in NR3 format
See Also
•
"Introduction to :EXTernal Trigger Commands"
on page 331
•
":EXTernal:PROBe"
on page 333
•
"Introduction to :TRIGger Commands"
on page 935
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...