Obsolete and Discontinued Commands
36
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
1119
:CHANnel2:SKEW
(see
page 1208
)
Command Syntax
:CHANnel2:SKEW <skew value>
<skew value> ::= skew time in NR3 format
<skew value> ::= -100 ns to +100 ns
The :CHANnel2:SKEW command sets the skew between channels 1 and 2.
The maximum skew is +/- 100 ns. You can use the oscilloscope's analog
probe skew control to remove cable delay errors between channel 1 and
channel 2.
Query Syntax
:CHANnel2:SKEW?
The :CHANnel2:SKEW? query returns the current probe skew setting for
the selected channel.
Return Format
<skew value><NL>
<skew value> ::= skew value in NR3 format
See Also
•
"Introduction to :CHANnel<n> Commands"
on page 264
N O T E
The :CHANnel2:SKEW command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :CHANnel<n>:PROBe:SKEW command (see
page 276
)
instead.
N O T E
This command is only valid for the two channel oscilloscope models.
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...