436
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
21
:MEASure Commands
:MEASure:DUAL:VPP
(see
page 1208
)
Overview
This measurement is available with the N2820A high sensitivity current
probe when both the Primary and Secondary probe cables are used. This
measurement joins the Zoom In waveform data below the probe's clamp
level with Zoom Out waveform data above the probe's clamp level to create
the waveform on which the measurement is made.
Command Syntax
:MEASure:DUAL:VPP [<source1>][,<source2>]
<source1>,<source2> ::= CHANnel<n> with N2820A probe connected
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:DUAL:VPP command installs a screen measurement and
starts a vertical peak- to- peak measurement.
If the optional source parameter(s) are specified, the currently specified
source(s) are modified.
Query Syntax
:MEASure:DUAL:VPP? [<source1>][,<source2>]
The :MEASure:DUAL:VPP? query measures the maximum and minimum
vertical value for the selected source, then calculates the vertical
peak- to- peak value and returns that value. The peak- to- peak value (Vpp)
is calculated with the following formula:
Vpp = Vmax - Vmin
Vmax and Vmin are the vertical maximum and minimum values present on
the selected source.
Return Format
<value><NL>
<value> ::= vertical peak to peak value in NR3 format
See Also
•
":MEASure:DUAL:CHARge"
on page 432
•
":MEASure:DUAL:VAMPlitude"
on page 433
•
":MEASure:DUAL:VAVerage"
on page 434
•
":MEASure:DUAL:VBASe"
on page 435
•
":MEASure:DUAL:VRMS"
on page 437
•
"Introduction to :MEASure Commands"
on page 420
•
":MEASure:SOURce"
on page 458
•
":MEASure:VMAX"
on page 474
•
":MEASure:VMIN"
on page 475
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...