:TRIGger Commands
32
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
983
:TRIGger:PATTern:QUALifier
(see
page 1208
)
Command Syntax
:TRIGger:PATTern:QUALifier <qualifier>
<qualifier> ::= {ENTered | GREaterthan | LESSthan | INRange | OUTRange
| TIMeout}
The :TRIGger:PATTern:QUALifier command qualifies when the trigger
occurs:
•
ENTered — when the pattern is entered.
•
LESSthan — when the pattern is present for less than a time value.
•
GREaterthan — when the pattern is present for greater than a time
value. The trigger occurs when the pattern exits (not when the
GREaterthan time value is exceeded).
•
TIMeout — when the pattern is present for greater than a time value. In
this case, the trigger occurs when the GREaterthan time value is
exceeded (not when the pattern exits).
•
INRange — when the pattern is present for a time within a range of
values.
•
OUTRange — when the pattern is present for a time outside of range of
values.
Pattern durations are evaluated using a timer. The timer starts on the last
edge that makes the pattern (logical AND) true. Except when the TIMeout
qualifier is selected, the trigger occurs on the first edge that makes the
pattern false, provided the time qualifier criteria has been met.
Set the GREaterthan qualifier value with the
:TRIGger:PATTern:GREaterthan command.
Set the LESSthan qualifier value with the :TRIGger:PATTern:LESSthan
command.
Set the INRange and OUTRange qualifier values with the
:TRIGger:PATTern:RANGe command.
Set the TIMeout qualifier value with the :TRIGger:PATTern:GREaterthan
command.
Query Syntax
:TRIGger:PATTern:QUALifier?
The :TRIGger:PATTern:QUALifier? query returns the trigger duration
qualifier.
Return Format
<qualifier><NL>
See Also
•
"Introduction to :TRIGger Commands"
on page 935
•
":TRIGger:PATTern:GREaterthan"
on page 981
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...