:WAVeform Commands
Data Acquisition Types
There are four types of waveform acquisitions that can be selected for
analog channels with the :ACQuire:TYPE command (see
page 240
):
NORMal, AVERage, PEAK, and HRESolution. Digital channels are always
acquired using NORMal. When the data is acquired using the :DIGitize
command (see
page 199
) or :RUN command (see
page 220
), the data is
placed in the channel buffer of the specified source.
Once you have acquired data with the :DIGitize command, the instrument
is stopped. If the instrument is restarted (via the programming interface
or the front panel), or if any instrument setting is changed, the data
acquired with the :DIGitize command may be overwritten.You should first
acquire the data with the :DIGitize command, then immediately read the
data with the :WAVeform:DATA? query (see
page 1029
) before changing
any instrument setup.
A waveform record consists of either all of the acquired points or a subset
of the acquired points. The number of points acquired may be queried
using :ACQuire:POINts? (see
page 232
).
Helpful Hints:
The number of points transferred to the computer is controlled using the
:WAVeform:POINts command (see
page 1032
). If :WAVeform:POINts
MAXimum is specified and the instrument is not running (stopped), all of
the points that are displayed are transferred. This can be as many as
4,000,000 in some operating modes or as many as 8,000,000 for a digital
channel on the mixed signal oscilloscope. Fewer points may be specified to
speed data transfers and minimize controller analysis time. The
:WAVeform:POINts may be varied even after data on a channel is acquired.
However, this decimation may result in lost pulses and transitions. The
number of points selected for transfer using :WAVeform:POINts must be an
even divisor of 1,000 or be set to MAXimum. :WAVeform:POINts determines
the increment between time buckets that will be transferred. If POINts =
MAXimum, the data cannot be decimated. For example:
•
:WAVeform:POINts 1000
— returns time buckets 0, 1, 2, 3, 4 ,.., 999.
•
:WAVeform:POINts 500
— returns time buckets 0, 2, 4, 6, 8 ,.., 998.
•
:WAVeform:POINts 250
— returns time buckets 0, 4, 8, 12, 16 ,.., 996.
•
:WAVeform:POINts 100
— returns time buckets 0, 10, 20, 30, 40 ,..,
990.
Analog Channel Data
NORMal Data
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...