716
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
28
:SBUS<n> Commands
:SBUS<n>:I2S:TRIGger:PATTern:DATA
(see
page 1208
)
Command Syntax
:SBUS<n>:I2S:TRIGger:PATTern:DATA <string>
<string> ::= "n" where n ::= 32-bit integer in signed decimal when
<base> = DECimal
<string> ::= "nn...n" where n ::= {0 | 1 | X | $} when
<base> = BINary
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F | X | $} when
<base> = HEX
The :SBUS<n>:I2S:TRIGger:PATTern:DATA command specifies the I2S
trigger data pattern searched for in each I2S message.
Set a <string> bit to "0" or "1" to set the corresponding bit in the data
pattern to low or high, respectively.
Set a <string> bit to "X" to ignore (mask off) that bit in the data pattern.
Use the "$" character to indicate that the value of the corresponding bit
will not be changed (the existing bit value is used).
When <base> = DECimal, the "X" and "$" characters cannot be entered.
When queried, the "$" character is returned when any bits in the pattern
have the value of "X" and <base> = DECimal. When any bits in a given
nibble have the value of "X" and <base> = HEX, the "$" character is
returned for the corresponding nibble.
N O T E
<base> is specified with the :SBUS<n>:I2S:TRIGger:PATTern:FORMat command. The
default <base> is DECimal.
N O T E
The :SBUS<n>:I2S:TRIGger:PATTern:DATA command specifies the I2S trigger data pattern
used by the EQUal, NOTequal, GREaterthan, and LESSthan trigger conditions. If the
GREaterthan or LESSthan trigger condition is selected, the bits specified to be masked off
("X") will be interpreted as 0's.
N O T E
The length of the trigger data value is determined by the :SBUS<n>:I2S:RWIDth and
:SBUS<n>:I2S:TWIDth commands. When the receiver word size is less than the transmitter
word size, the data length is equal to the receiver word size. When the receiver word size is
greater than the transmitter word size, the data length is equal to the transmitter word size.
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...