492
Agilent InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
22
:MEASure Power Commands
:MEASure:CRESt
(see
page 1208
)
Command Syntax
:MEASure:CRESt [<source>]
<source> ::= {CHANnel<n>| FUNCtion<m> | MATH<m>}
<n> ::= 1 to (# analog channels) in NR1 format
<m> ::= 1-4 in NR1 format
The :MEASure:CRESt command installs a crest factor measurement on
screen.
The <source> parameter is the channel probing current or voltage. This
source can also be specified by the :MEASure:SOURce command.
Crest factor is a measure of power quality. It is the ratio between the
instantaneous peak AC line current (or voltage) required by the load and
the RMS current (or voltage). For example: Ipeak / IRMS or Vpeak /
VRMS.
Query Syntax
:MEASure:CRESt? [<source>]
The :MEASure:CRESt query returns the measured crest factor.
Return Format
<return_value><NL>
<return_value> ::= the crest factor value in NR3 format
See Also
•
":MEASure:SOURce"
on page 458
•
":POWer:QUALity:APPLy"
on page 579
Summary of Contents for InfiniiVision 4000 X-Series
Page 1: ...s1 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 32: ...32 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide ...
Page 52: ...52 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 2 Setting Up ...
Page 66: ...66 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 3 Getting Started ...
Page 242: ...242 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 7 ACQuire Commands ...
Page 252: ...252 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 8 BUS n Commands ...
Page 300: ...300 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 12 DEMO Commands ...
Page 330: ...330 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 15 DVM Commands ...
Page 386: ...386 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 18 HARDcopy Commands ...
Page 484: ...484 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 21 MEASure Commands ...
Page 1098: ...1098 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 34 WGEN w Commands ...
Page 1174: ...1174 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide 37 Error Messages ...
Page 1352: ...1352 Agilent InfiniiVision 4000 X Series Oscilloscopes Programmer s Guide Index ...