DD+DIS155.11E
Repair and Service
Tools and auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 30 of 31
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential
3
Remote Service
Purpose
Remote connection to the digitizer means:
Connection via a PC in the local network or via internet.
To prevent that service actions initiated via service user interface interfere with the
operation by the customer, two safety levels are built in the software:
Level 1: When executing the function (e.g. software installation), the digitizer status
indicator changes to red constant. The user cannot scan an image.
Note, that the status indicator only changes to red, when the function is really
executed (e.g. the user presses OK). Just by opening the menu, the safety level 1
is not yet activated.
Level 2: The user has to confirm in the error viewer (installed on the NX
workstation), that the function can be executed.
The table below shows the safety level per service menu:
Service Menu Name
Reference
Level 1:
Scanning blocked
Level 2:
Confirmation
required
Logout
no no
Device Info
no no
Network Info
no no
Info Counter
no no
Error Statistics
no no
Service Activity Logfile
no no
Software Installation
yes
no
Digitizer Settings
yes
no
Destination Settings
yes
no
Site Specific Data
no no
Installation Wizard
yes
no
Error Viewer Installation
no* no
Log Settings
no no
Error Code Explanation
no no
Device Test
yes
no
Diagnose Cycle
yes yes
Endurance Run Cycle
yes yes