DD+DIS155.11E
Repair and Service
Tools and auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2.5
Diagnose & Repair
2.5.1
Error Code Explanation
Purpose
The menu “Explain Error Code” list the description for the digitizer error codes which
are reported by the error viewer.
Usage
(1) Select service menu:
<Explain
error code>
(2) Enter the error code and click button:
<
Get Error Description
>
The error description is displayed.
Figure 15
NOTE:
The digitizer issues two types of error codes:
Error Code Type
Example
Example code to be entered in
the Service User Interface
Pure service message
ARC1001 (25613)
ARC1001*
R&D internal error code
24588
24588
*Also the internal error code (in this example "25613") can be looked up.
Edition 1, Revision 1
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 20 of 31
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential