Repair and Service
Tools and auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2
Digitizer Service User Interface
Access to the “Service User Interface” for
the digitizer is available via browser
(Internet Explorer or Firefox).
Any PC with network access to the
digitizer can be used.
Figure 2: Digitizer Service User Interface
It is recommended using the NX workstation to access the digitizer service user
interface, as the NX workstation has the error viewer installed by default.
When using a different PC in the network, following service functions are not
accessible, as they require confirmation in the error viewer:
Diagnose
Cycle
Endurance Run Cycle
Shading
Calibration
For more information se also section 3, Remote Service.
Proceeding to access the Service User Interface:
(1) Start the browser via shortcut or by entering “iexplore” in the search window of
the start menu.
(2) Enter the digitizer IP address (e.g. 192.192.192.192) in the browser
address window. Note, that IP address 192.192.192.192 does not work with
Internet Explorer 9 and software < ARC_1404. See also CR 10-X SB08.
(3) Enter username and password which is provided during the web based training.
Proceeding to quit the Service User Interface:
(1) Select:
<Logout>
(2) Follow the Logout wizard.
(3) Close the browser window.
Edition 1, Revision 1
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 6 of 31
03-2013
Type 5151 / 100 / 110 / 200 / 210