DD+DIS155.11E
Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.8 / Page 8 of 13
04-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential
4
Software Release ARC_1302
Release Date
09/2012
Digitizer SNs
Software ARC_1302 is installed in digitizers with
SN
≥
1517 and additionally in SN 1381, 1424, 1428,
1429, 1430, 1433, 1435
Availability
Agfa HealthCare Library*
USB flash drive, delivered with the digitizer
Prerequisites
The NX workstation connected to the digitizer requires
software
≥
NX 2.0.8700 / 3.0.8700.
For additional system component compatibility
information refer to following documents:
Solution Structure Data Document, Document ID*
30781759 (
CR/DR Interoperability Matrix, Document ID*
31333326 (
Reference Document
DD+DIS189.12E, Service Bulletin No. 04,
Software Update ARC_1302 available (
/
* On the Agfa HealthCare Library only the latest software release can be downloaded.
4.1
Solved Software Issues in Software ARC_1302
Solution for Headquarter Complaint HQ_1207130004:
With ARC_1302 also a new version of the digitizer operating system is installed. This
solves the issue with some digitizer USB flash drives, being too slow for software
installation.
In software ARC_1201 a workaround was implemented to prevent a software crash in
case of Photomultiplier flash memory errors. Now the final solution is implemented, to
prevent read and write errors of the Photomultiplier flash memory.
Solved sporadic software issue with error viewer being offline each 15 seconds. The
issue may appear if digitizer and PC are switched on at the same time.
Solved issue with green blinking status LED after scan for several minutes, if an error
occurred during scan.
The feedback messages during service function “endurance run” are now updated with
each step (feeding IP, scanning, erasing etc.). Before, the feedback messages stopped
with “feeding IP”, despite endurance run worked fine.