2.5.6
Reset Relative Counters
Purpose
With the menu “Reset Relative Counters” the relative counters* of the info counter
are cleared.
This may be required to easily track new info counter entries after a service
intervention.
* Number of test cycles per format; number of start-ups and resets; Retries; Incorrect User Handling;
Cassette and IP jams; Error List; Warning List
Usage
(1) Select service menu:
<Reset
Relative Counters>
(2) Click:
<OK>
Figure 22
2.5.7
Clear Image Queue
Purpose
There are cases, where an image cannot be transmitted, and stays in the queue
forever, despite the processing station software is running. This may happen during
setup of the digitizer. Examples:
An image is scanned while the network is disconnected. At the same time the
destination settings are changed (e.g. IP address of NX).
Destination AE title is wrong configured in the digitizer.
In this case after each subsequent scan a “destination not available” or “destination is
invalid” warning is displayed.
With the menu “Clear Image Queue” all images in the image queue can be erased.
WARNING:
By clearing the image queue all images in the queue of the digitizer are erased.
Retake possible.
Before clearing the image queue, check all other troubleshooting options to get the
image transmitted to the destination.
Edition 1, Revision 1
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 25 of 31
03-2013
Type 5151 / 100 / 110 / 200 / 210