DD+DIS170.11E
Installation Planning
1
2
3
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 11 / Page 20 of 24
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential
8.4
Functional Data of the optional UPS
Rated voltage [~V] UPS Powerware
5115-750 (ABC Code: EGPSE)*
120 = Default
110 = User adjustable via rear panel
DIP switches
Rated voltage [~V] UPS Powerware
5115-750i (ABC Code: EGPTG)*
230 = Default
220 or 240 = User adjustable via rear
panel DIP switch
Frequency [Hz]
50/60 (autosensing)
Battery runtime [min]
> 5
Output Power [W]
750
Dimensions (HxWxD) [mm/in]
192 x 150 x 270 / 7.6 x 5.9 x 10.6
Weight [kg/lb]
7.8 / 17.2
More info
Refer to www.powerware.com
8.5
Heat Emission
Heat emission type 5151 / 100
and 110 [W / BTU per hour]
Standby:
Average during
scan:
41 / 140
65 / 222
Heat emission type 5151 / 200
and 210 [W / BTU per hour]
Standby:
Average during
scan:
22 / 75
78 / 266
8.6
Acoustic Noise Level
Acoustic Noise Level
according ISO 7779
During scanning:
65 dB(A) maximum
8.7
Boot-up Time and Warm-up Time
Boot-up time and Warm-up time
Approximately 3 minutes
Warm-up time “cold start”
Image quality within specification
after boot-up time