Repair and Service
Tools and auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
(3) Put a cassette into the digitizer.
(4) Press the grey marked buttons to
perform the required actions.
See also NOTE below.
Figure 18
IMPORTANT:
The Error Viewer and related confirm / cancel window only pops up at the
NX workstation. This means, that in any case entering the diagnose cycle
window needs to be confirmed at the NX workstation.
Only leave the Diagnose Cycle window by the
CANCEL
button. Leaving it via
browser back button requires a reboot of the digitizer to allow further service
activities.
NOTE:
Description of the colored rectangles shown in the diagnose cycle overview:
A light blue displayed rectangle shows other statuses of the complete
cassette cycle.
A green displayed rectangle shows the current status.
A dark blue displayed rectangle shows the next possible action.
A gray displayed rectangle shows the next possible action.
Edition 1, Revision 1
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 22 of 31
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential