DD+DIS155.11E
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
5.3
Performing Setup of the NX Workstation
NOTE:
The NX workstation setup comprises following steps:
Performing the operating system settings (date & time, network, language)
Installing the license file
Activating the NX software
Configuring the digitizer on the NX workstation
(1) Switch on the digitizer.
The red status LED starts flashing.
(2) Switch on the NX Workstation.
(3) Login with service user name and
password obtained in the NX
classroom training.
install_poster_04.cdr
User name
Password
Figure 19
IMPORTANT:
With direct network connection observe
the correct switch on sequence: First
switch on digitizer, then the NX
workstation.
To avoid a fixed switch on sequence
with direct network connection, use one
of the workarounds mentioned on
page 21.
install_poster_003a.cdr
Direct network connection:
Observe switch-on sequence
or use one of the available workarounds.
Figure 20
Edition 1, Revision 4 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Chapter 1 / Page 23 of 47
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential