DD+DIS155.11E
Repair and Service
Tools and auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2.4.2
Digitizer Settings
Purpose
The menu “Digitizer Settings” allows entering data to modify:
The local settings of the digitizer:
o
Date and Time
o
User interface language (language of the user messages displayed in the
error viewer on the NX workstation)
The digitizer network settings.
Usage
(1) Select service menu:
<Digitizer
Settings>
.
(2) Perform the required settings. See
also NOTE below.
(3) If you want to go back to the
default digitizer settings: Select
<Load default settings>
.
(4) Select
<Load Digitizer
Network Data from USB>
to
read out the data from an adc.cpf
file on the digitizer USB flash drive.
This CPF file:
Can be created with the CCM
tool* or
Can be exported from the NX
configuration tool.
Figure 10
*The CCM tool (
C
onfiguration and
C
ustomization
M
anager) is by default installed on the NX workstation.
NOTE:
The configuration of the digitizer settings is described in the digitizer Service
Manual, chapter 1, Controls, Connections and Setup Procedures.
Function “Load default settings” is available as of software version ARC_1404
only.
Edition 1, Revision 1
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 14 of 31
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential