DD+DIS155.11E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
(8) A screen with instructions is
displayed.
Enter the exposed cassette in the
digitizer: Scanning starts.
Figure 4
(9) Wait until scanning is finished.
A static red status LED indicates
scanning.
A static green status LED indicates
that scanning is finished: The
shading calibration curve is
displayed.
Figure 5
IMPORTANT:
If a message pops up which indicates a failed shading calibration (e.g. “Underexposed
- too low pixel values …) refer to section 2.5 for possible remedies.
Edition 1, Revision 1
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 3.6 / Page 8 of 10
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential