
DD+DIS155.11E
Repair and Service
Troubleshooting
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Chapter 3.3 / 19 of 46
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential
CODE
Message
ARC1203
Clean Up procedure failed.
Meaning: The IP is not transported into the cassette successfully.
Reason 1: Handling of the IP not successful.
Cure 1: Restart Digitizer.
Reason 2: Cassette not inserted straight into 'Cassette Unit'.
The IP has been lost at Slow Scan Drum and is placed at IP-Tray.
Cure 2: Check the Cassette for straight entry.
The distances of the arrow marks at tube side of he inserted Cassette have to
be equal to the slot. In case of repeated appearance of this failure the
'Cassette Unit' has to be replaced.
Open the Digitizer and remove the lost IP.
Reason 3: IP has jammed.
Cure 3: Open Digitizer and remove jammed IP.
Reason 4: The IP is not transported straight through the Digitizer.
The IP has been lost at Slow Scan Drum and is placed at IP-Tray.
Cure 4: Open the Digitizer and remove the lost IP.
Check by help of Diagnose Cycle the 'Cassette Unit' for feeding the IP straight
into the Digitizer. In case of repeated appearance of this failure the 'Cassette
Unit' has to be replaced.