DD+DIS155.11E
Repair and Service
Tools and auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2.4.7
Log Settings
Purpose
The menu “Log Settings” allows to change the default log settings. This may be
required to trace sporadic problems.
Usage
(1) Select service menu:
<Log
Settings>
(2) Adapt the required settings:
Log Level:
Standard or Extended.
Default: Standard.
Hours until extended Log Level is
automatically reset to standard.
Default: 2
(3) Click:
<OK>
Figure 14
NOTE:
Only use the "extended logging" function if really required for troubleshooting:
Extended logging reduces the performance of the system.
Edition 1, Revision 1
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 19 of 31
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential