CODE
Message
ARC2504
Motor error at the Slow Scan Unit.
Meaning: Motor selftest of the 'Slow Scan Motor' detects an error.
Reason 1: The cables are defective or disconnected.
Cure 1: Check cables and connectors between 'Power Board' and 'Slow Scan
Motor'.
Replace 'Y-Cable Power Board to Motors'.
Reason 2: The 'Slow Scan Motor' is defective.
Cure 2: Replace the 'Slow Scan Motor'.
Reason 3: 'Power Board' defective.
Cure 3: Replace 'Power Board'.
ARC2600
Erasure Unit reports supply voltage error.
Meaning: The fuse for supply voltage to the 'Erasure Unit' is defective. 'Power
Board' defective.
Reason 1: Defective 'Power Board' caused by short circuit at the 'Cable Power
Board to Erasure Unit'.
Cure 1: Replace 'Power Board' AND 'Cable Power Board to Erasure Unit'
together!
Reason 2: Defective 'Power Board' caused by short circuit at the 'Erasure Unit'.
Cure 2: Replace 'Power Board' AND 'Erasure Unit' together!
ARC2601
Erasure Unit reports LED current error.
Meaning: The Erasure Unit could not be switched on successfully. LED
currents are not OK.
Reason 1: 'Cable Power Board to Erasure Unit' defective or disconnected.
Cure 1: Check cables and connectors to the 'Erasure Unit'.
Replace 'Cable Power Board to Erasure Unit'.
Reason 2: 'Erasure Unit' current path broken or partly disconnected.
A LED line is defective. Only one colour 'on' during selftest.
Cure 2: Check the short cable and connector to the overtemperature protection
at connector P2.
Let the 'Erasure Unit' cool down if hot. Replace 'Erasure Unit'.
Reason 3: 'Power Board' current path interrupted or electronic driver defect.