Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
(65) Enter the data in the
Device Configuration
Digitizer
Emergency Procedure
window according to the customer needs.
The following table just shows examples:
#
Parameter
Value
1 Age
Group:
17+
2
Exam Group:
Carpus (Knee)
3
Exposure Type:
DLPMO Left
Following two steps are for CR 12-X only:
(66) Select in the NX Configuration Tool Menu:
User Interface
Identification Configuration
Edit
Image Details
(67) In dropdown box
Content
, add label
Scan Resolution
.
Figure 54
NOTE:
With scan resolution parameter enabled in the NX “Edit Image Detail” window, the user
can switch scan resolution before scanning.
The default scan resolution settings are defined in the digitizer configuration window.
Recommendation: Leave settings on default.
For more information see also Application Note NX 2.0.8800 / 3.0.8800,
Document ID 39248556.
Edition 1, Revision 4 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Chapter 1 / Page 39 of 47