DD+DIS170.11E
Installation Planning
1
2
3
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 11 / Page 13 of 24
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential
5
Ambient Conditions during Transport, Storage and Operation
5.1
Climatic Conditions
Climatic
Classification
Related
Temperature
Values
Related
Humidity
Values
Transport in package
2K2 according
IEC 721-3-2
-25° C to + 55°C
-13° F to + 140° F
0 % to 75 %
Storage in package
1K4 according
IEC 721-3-1
-25° C to + 55°C
-13° F to + 131° F
10 % to 100 %
Operation 3K2
according
IEC 721-3-3
+ 15° C to + 35°C
+59° F to + 95° F
10 % to 75 %
Transport in mobile X-
Ray room
5K1 according
IEC 721-3-5
+5° C to + 40° C
+41° F to + 140° F
10 % to 75 %
Operation in mobile X-
Ray room
3K2 according
IEC 721-3-3
+15 ° C to + 35°C
-59° F to + 95° F
10 % to 75 %
5.2
Mechanical Conditions
Mechanical Classification
Transport in package
2M3 according IEC 721-3-2
Storage in package
1M2 according IEC 721-3-1
Operation
Transport in truck for mobile X-Ray
5M3 according IEC 721-3-5*
Operation in truck for mobile X-Ray
* if digitizer is mounted on the optional Mobile Kit
5.3
Light Conditions
The digitizer may not be operated in direct sun light. No influence on image quality is
guaranteed if operated in ambient light of maximum 2500 Lux.