Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.8 / Page 6 of 13
04-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential
Photomultiplier flash memory errors are repaired automatically: The software
doesn't crash anymore if the Photomultiplier flash memory is corrupted.
Galvo adjustment is optimized: The adjustment is checked and repeated once per
minute if required.
Infocounter values like installation/manufacture date and machine running time are
corrected.
Image is not lost anymore for following special case:
- Network is disconnected during scan and
- Cassette is not removed after scan and
- Digitizer is switched off afterwards
2.2
Known Software Issues of Software ARC_1201
When running service menu “Endurance run” without scanning or without erasing: The
status is not updated. It remains on "feeding IP".