DD+DIS155.11E
Repair and Service
Troubleshooting
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Chapter 3.3 / 26 of 46
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential
CODE
Message
ARC2201
Photomultiplier data is not accessible.
Meaning: The access to data stored on the 'PMT with Light Collector' flash
memory failed.
Reason 1: The cables to 'PMT with Light Collector' are defective or
disconnected.
Cure 1: Check cables and connectors between 'PMI Board' and 'PMT with
Light Collector'.
Replace 'Data Cable PMI Board to Tube Board'. Replace 'Signal Cable PMI
Board to Tube Board'.
Reason 2: The 'PMT with Light Collector' flash memory contains invalid data or
is empty.
Cure 2: Replace 'PMT with Light Collector'.
ARC2202
Photomultiplier operating conditions not adjustable.
Meaning: At the 'PMT with Light Collector' the setting failed.
Or the PMT supply voltage is out of specification.
Or the communication to the PMT is not successful.
Reason 1: The cables to 'PMT with Light Collector' are defective or
disconnected.
Cure 1: Restart Digitizer to get relevant selftest result. Try to scan an IP again
to test the PMT.
Check cables and connectors between 'PMI Board' and 'PMT with Light
Collector'.
Replace 'Data Cable PMI Board to Tube Board'. Replace 'Signal Cable PMI
Board to Tube Board'.
Reason 2: The 'PMT with Light Collector' is defective.
Cure 2: Replace 'PMT with Light Collector'.
Reason 3: 'PMI Board' defective.
Cure 3: Replace 'PMI Board'.