Power Quality Analyzer Model 8335
27
Max intrinsic error of the flicker severity measurement (Pst)
120 V lamp
60 Hz network
230 V lamp
50 Hz network
Rectangular variations
per minute
(50% duty cycle)
2
Pst
∈
[1;4]
± 5%
Pst
∈
[1;4]
± 5%
7
Pst
∈
[1;7]
± 5%
Pst
∈
[1;4]
± 5%
39
Pst
∈
[1;12]
± 5%
Pst
∈
[1;10]
± 5%
110
Pst
∈
[1;12]
± 5%
Pst
∈
[1;10]
± 5%
1620
Pst
∈
[1;12]
± 15%
Pst
∈
[1;10]
± 15%
Ratio
Minimum
Maximum
Voltage
100
1000 x
√
3
9,999,900 x
√
3
0.1
Current
(1)
1
60,000 / 1
(1) Only for the MN193 clamp (5 A) and the 5 A adapter.
Measurement
Measurement Range
Minimum
with minimum ratio(s)
Maximum
with maximum ratio(s)
RMS
&
RMS ½
Voltage
Simple
58 mV
207.8 GV
Compound
58 mV
415.7 GV
Direct
Voltage
(DC) &
Peak
Voltage
Simple
58 mV
293.9 GV
Compound
58 mV
587.9 GV
RMS & RMS ½ Current
5 mA
360.0 kA
Peak Current
5 mA
509.1 kA
Active Power
0.289 mW
74.82 PW
Reactive Power
0.289 mVAR
74.82 PVAR
Apparent Power
0.289 mVA
74.82 PVA
Active Energy
1 mWh
9,999,999 EWh
(1)
Reactive Energy
1 mVARh
9,999,999 EVARh
(1)
Apparent Energy
1 mVAh
9,999,999 EVAh
(1)
(1) The energy corresponds to more than 15,000 years of the associated maximum power (max ratios).