Power Quality Analyzer Model 8335
7.2 Recharging the Battery
Battery life is 10 hours when the battery is completely charged and no recording is
in progress. Battery life is approximately 35 hours when recording and the display
is off.
The batteries automatically begin recharging when the PowerPad
®
is connected
to the AC power supply. Only use the supplied AC power supply to recharge the
batteries of the PowerPad
®
. It will take about 5 hours to fully charge completely
drained batteries. The batteries will not be depleted when the unit is connected to
the power supply.
The instrument will not recharge if the message “Instrument will soon
turn OFF” is displayed on the screen. The Enter button must be pressed
or the instrument must be turned OFF before recharging will start.
7.3 Changing the Battery
WARNING:
•
When changing the battery, disconnect all instrument inputs and turn the
equipment off. There must be a delay of at least one minute without the
battery being connected.
• Do not expose the battery to heat exceeding 212°F (100°C)
• Do not short-circuit the battery terminals
To change the battery, disconnect from any input and make sure the instrument is
turned off.
With a coin, remove back panel and replace with a 9.6V NiMH factory supplied
battery pack.
NOTE:
The memory card is accessible when the battery is removed from the com-
partment.
7.4 Cleaning
Disconnect the instrument from any source of electricity.
• Use a soft cloth, lightly dampened with soapy water
• Wipe with a damp cloth and then dry with a dry cloth
• Do not splash water directly on the clamp
• Do not use alcohol, solvents or hydrocarbons