CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
TROUBLESHOOTING
1.2 ERROR MESSAGES
1 - 6
1
Alarm 4 (s)
Sample Tube not sensed correctly.
Alarm 4 (t)
Safety Cover removed during sample processing.
Alarm 4 (u)
Rack Number Misread. Barcode reader did not properly read a
rack barcode.
Autosampler Command Negatively Acknowledged
Reseat
RS232
Cable At
Both
Ends
And
Restart
Auto Stand-by Mode
None
Blood Detected
in Sample Line
If system automatically rinses, and error does not recur.
None
If "Blood Detected in Sample Line" error displays on bulletin line.
If FATAL error displays on bulletin line.
Cycle
Power,
then
Check Rack Alarm
None
Cover Open
Table 1 - 1 Error Messages(continued)
MESSAGE
CORREC-
TIVE
ACTION