TROUBLESHOOTING IP - 1 Blood Detected In Sample Line, 1 of 3
1 - 10
1
IP - 1 Blood Detected In Sample Line, 1 of 3
B lo o d in
L in e
?
C h e c k fo r B lo o d in
S a m p le L in e N e a r
B lo o d In -L in e
S e n s o r
N o
S T A R T
F ro m th e D a ta S ta tio n M a in M e n u
-
G o to S p e c ia l P ro to c o ls a n d P re ss
C L E A N N E E D L E K e y
-
C h e c k A ll T u b in g , T -C o n n e c to rs , a n d
Y -F ittin g s le a d in g T h ro u g h S O L # 9 6
a n d # 9 2 E n d in g a t W a s te C h a m b e r
# 2
-
C h e c k 1 2 " (8 2 .7 k P a ) V a c u u m U se d
fo r E v a c u a tin g L e fto v e r S a m p le
-
C h e c k D ilu e n t L in e le a d in g T h ro u g h
S O L # 9 1 a lo n g W ith 4 P S I (2 7 .6 kP a )
U s e d to F lu sh th e L in e
-
C h e c k V a c u u m C o n n e c tio n fro m
W a s te C h a m b e r # 2 to S a m p le
L o a d e r W a sh B lo c k
Y E S