CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 5.2 Tube Detect Sensor (Beam Block) Adjustment
3 - 74
3
Figure 3-34 Beam Block Test Points - [NON-CE MARK]
d. With No Racks on Loader, turn VR8 fully counter-clockwise and verify that voltage (volt meter
on AC) drops below .010 VAC.
SL-186D
TP22
TP23
VR8
CN5
Measure
R12 Here
Measure
C37 Here
[OLD CPU MAIN BOARD]