CELL-DYN
®
3/35 Sample Loader TS Guide
9140234B-June 1997
VERIFICATION PROCEDURESVP - 6.4 Mixing Contact Sensor Check/Adjustment
3 - 96
3
Figure 3-49
6. Verify Contact Sensor clearance.
[
CE MARK] Clearance Check is NOT required.
[NON-
CE MARK] Mixing Paddle must clear Sensor by .010-015in. (0.25-0.38mm.); .015in.
(0.38mm) preferred (
SL-271
Flush With
Bottom of Sensor
All G Loaders
G Loader