3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURESVP - 15.1 Barcode Reader Configuration (Europe Only) 3 - 164
3
7. Insert Service Disk into Data Station floppy disk drive.
8. Turn Sample Loader and Data Station power ON.
9. At A:> prompt, type VBCDIAG and press [ENTER].
10.Data Station displays:
• Function Key Information
• RS-232 Configuration
CAUTION:If the message “PORT 1: UNABLE TO GO ONLINE ... COMM STATUS” displays,
repeat
NOTE:If procedure fails, check cable wiring according to
. If problem
persists, perform
11.Change Baud Rate to 1200.
a. Press [F3] and wait 20 seconds.
b. Verify 1200 Baud Rate is displayed.
c. To keep existing Barcode settings, continue with Step 12.
d. To change Barcode settings to default values, press [F3] again.
System beeps.
12.Edit Barcode parameters.
NOTE:System responds slowly. Observe screen while entering commands and confirm
acceptance by system.